Determination of Sr Content in Mg-Sr Alloys Using Faas

Article Preview

Abstract:

The Air-acetylene Flame Atomic Absorption Spectrometry (FAAS) was used to measure Sr content in Mg-Sr Alloys. The experiment conditions including acetylene flow rate, measurement height, flame wavelength were studied and the disturbing factors were discussed. It was found that, as dissolution agent of alloys and LaCl3 as the deionization agent and flame wavelength being 460.7nm, the detection limits is 1.16mg/L and the relative standard deviations is less than 0.3% and the recovery rates of Sr is more than 95%. The advantages of this method were summarized as: high sensitivity and precision, low chemistry disruption, good selectivity, good recurrence characteristics of the measuring conditions. The method is suitable for the control analysis and sample-system analysis of determination Sr content in the Mg-Sr Master Alloys.

You might also be interested in these eBooks

Info:

Periodical:

Materials Science Forum (Volumes 561-565)

Pages:

215-218

Citation:

Online since:

October 2007

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2007 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] Bai Jing,Sun Yangshan,Xue Feng,Xue Shan,Qiang Jing,Tao Weijian. Effect of extrusion on microstructures and mechanical and creep properties of Mg-Al-Sr and Mg-Al-Sr-Ca alloys. Scripta Materialia[J],55(2006) 1163-1166.

DOI: 10.1016/j.scriptamat.2006.08.020

Google Scholar

[2] Deng Bo. The Principle, Techniques and Application of Atomic Absorption Spectrometry [M]. Beijing: Tsinghua University Press,(2004).

Google Scholar

[3] HU Junyong, XU Guang, WANG Ping. Determination of Strontium in Aluminium-strontium Master Alloy by Atomic Absorption Spetrophotometry [J]. Jiangxi Metallurgy, 2003,23(5):36-37.

Google Scholar