Metallosis - Literature Review and Particular Cases Presentation

Article Preview

Abstract:

Metallosis is an adverse event developed in relation with an orthopedic implant. It was brought into attention by metal on metal total hip arthroplasty. Lately, cases were diagnosed in total knee, total elbow, and spinal surgery. Metallic debris - released because of wear or corrosion - start an inflammatory response in a chronic condition. Beside local effects, systemic effects are also described; among them toxic effects, neurological and psychiatric, alteration in thyroid and heart function, skin rushes and discoloration. Diagnosis is helped by x-ray examination but is based on fluid aspiration with ions level determination and histological examination. Osteolysis phenomena induced by metallosis may compromise bone ingrowth and promote implant loosening; as result bone stock may be compromised. The cases we present have a particular development pattern; each of them was initiated as a result of atypical behavior of the implants. Two of them necessitated bone grafting in order to replace the reduced bone stock and all three required revision surgery. The cases emphasize the diagnosis methodology and the possible complications encountered during orthopedic implant surgery.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

77-90

Citation:

Online since:

July 2017

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2017 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

* - Corresponding Author

[1] Korovessis P, Petsinis G, Repanti M, Repantis T., Metallosis after contemporary metal-on-metal total hip arthroplasty. Five to nine-year follow-up, J Bone Joint Surg Am, 88 (2006) 1183-1191.

DOI: 10.2106/jbjs.d.02916

Google Scholar

[2] Matsushita I., Morita Y., Gejo R., Kimura T., Severe bone defects and reduced mineralization caused by massive metallosis after total knee arthroplasty: histopathologic and bone morphometric findings, Mod Rheumatol, 17 (2007) 507-510.

DOI: 10.1007/s10165-007-0618-3

Google Scholar

[3] Hallab N., Merritt K., Jacobs J.J., Metal sensitivity in patients with orthopaedic implants, J Bone Joint Surg Am, 83 (2001) 428-436.

DOI: 10.2106/00004623-200103000-00017

Google Scholar

[4] Doorn P.F., Mirra J.M., Campbell P.A., Amstutz H.C., Tissue reaction to metal on metal total hip prostheses, Clin. Orthop, S (1996) S187–S205.

DOI: 10.1097/00003086-199608001-00017

Google Scholar

[5] Bensalah N., Pitting Corrosion, InTech (2012), 93-118.

Google Scholar

[6] Delaunay C., Petit I., Learmonth I.D., Oger P., Vendittoli P.A., Couple de frottement métal–métal des prothèses totales de hanches: surveillance du relargage des ions métalliques cobalt et chrome, Revue de Chirurgie Orthopédique et Traumatologique, 96 (2010).

DOI: 10.1016/j.rcot.2010.08.002

Google Scholar

[7] Pritchett J.M., Adverse reaction to metal debris: metallosis of the resurfaced hip, Current Orthopedic Practice, 23 (2012) 1-11.

DOI: 10.1097/bco.0b013e3182356075

Google Scholar

[8] Niculescu M., Laptoiu D., Miculescu F., Antoniac I., Metal allergy and other adverse reactions in patients with total hip replacement, Advanced Materials Research, 1114 (2015) 283-287.

DOI: 10.4028/www.scientific.net/amr.1114.283

Google Scholar

[9] Antoniac I., Handbook of Bioceramics and Biocomposites, Springer International Publishing, New York (2016) 935-967.

Google Scholar

[10] Atasiei T., Antoniac I., Laptoiu D., Failure causes in hip resurfacing arthroplasty – retrieval analysis, International Journal of Nano and Biomaterials, 3 (2011) 367-381.

DOI: 10.1504/ijnbm.2011.045882

Google Scholar

[11] Willert H.G., G.H. Buchhorn, A. Fayyazi, R. Flury, M. Windler, G. Koster, Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study, J Bone Joint Surg (Am), 87 (2005) 28–36.

DOI: 10.2106/jbjs.a.02039pp

Google Scholar

[12] Sansone V., Pagani D., Melato M., The effects on bone cells of metal ions released from orthopaedic implants. A review, Clinical Cases in Mineral and Bone Metabolism, 10 (2013) 34–40.

DOI: 10.11138/ccmbm/2013.10.1.034

Google Scholar

[13] Antoniac I., Laptoiu D., Miculescu F., Istrate R., Trisca-Rusu C., Microscopy analysis of total knee prosthesis failure caused by polyethylene wear, ECM, 16: S1 (2008) 54.

Google Scholar

[14] Cristescu I., Antoniac I., Vilcioiu D., Safta F., Analysis of centromedullary nailing with implant failure, KEM, 638 (2015) 130-134.

DOI: 10.4028/www.scientific.net/kem.638.130

Google Scholar

[15] Laurian T., Tudor A., Antoniac I., Miculescu F., A micro-scale abrasion test to study the influence of counterface roughness on the wear resistance of UHMWPE, JOAM, 9 (2007), 3383 – 3388.

Google Scholar

[16] Pandit H., Glyn-Jones S., McLardy-Smith P., Pseudotumours associated with metal-on-metal hip resurfacings, J Bone Joint Surg (Br), 90 (2008) 847–851.

DOI: 10.1302/0301-620x.90b7.20213

Google Scholar

[17] Visuri T.I., Pukkala E., Pulkkinen P., Paavolainen P., Cancer incidence and causes of death among total hip replacement patients: a review based on Nordic cohorts with a special emphasis on metal-on-metal bearings, Proc. ImechE J Eng Med, 220 (2006).

DOI: 10.1243/095441105x63282

Google Scholar

[18] Daley B., Doherty A.T., Fairman B., Case J.P., Wear debris from hip or knee replacement causes chromosomal damage in human cells in tissue culture, J Bone Joint Surg (Br), 86 (2004) 598–606.

DOI: 10.1302/0301-620x.86b4.14368

Google Scholar

[19] Hallab N.J., Jacobs J.J., Biologic effects of implant debris, Bull NYU Hosp Jt Dis., 67 (2009) 182-188.

Google Scholar

[20] Fehring T.K., Carter J.L., Fehring K.A., Odum S.M., Griffin W.L., Cobalt to Chromium Ratio is Not a Key Marker for Adverse Local Tissue Reaction (ALTR) in Metal on Metal Hips, J Arthroplasty, 30 (2015) 107-109.

DOI: 10.1016/j.arth.2015.02.049

Google Scholar

[21] Keegan G.M., Learmonth I.D., Case C.P., Orthopaedic metals and their potential toxicity in the arthroplasty patient. A review of current knowledge and future strategies, J Bone Joint Surg (Br), 89 (2007) 567–573.

DOI: 10.1302/0301-620x.89b5.18903

Google Scholar

[22] Lützner J., Hartmann A., Dinnebier G., Spornraft-Ragaller P., Hamann C., Kirschner S., Metal hypersensitivity and metal ion levels in patients with coated or uncoated total knee arthroplasty: a randomised controlled study, Int Orthop, 37 (2013).

DOI: 10.1007/s00264-013-2010-6

Google Scholar

[23] Tower Stephen S., Arthroprosthetic Cobaltism: Neurological and Cardiac Manifestations in Two Patients with Metal-on-Metal Arthroplasty: A Case Report, J Bone Joint Surg Am, 92 (2010) 2847-2851.

DOI: 10.2106/jbjs.j.00125

Google Scholar

[24] Oliveira C.A., Candelária I.S., Oliveira P.B., Figueiredo A., Metallosis: A diagnosis not only in patients with metal-on-metal prostheses, European Journal of Radiology Open, 2 (2015) 3–6.

DOI: 10.1016/j.ejro.2014.11.001

Google Scholar

[25] Post Z.D., Orozco F.R., Ong A.C., Metal sensitivity after TKA presenting with systemic dermatitis and hair loss, Orthopedics, 36 (2013) 525–528.

DOI: 10.3928/01477447-20130327-35

Google Scholar

[26] Callender V.D., Cardwell L.A., Munhutu M.N., Bigby U., Rodney I.J., Cutaneous metallosis in patient with knee prosthesis composed of cobalt-chromium molybdenum alloy and titanium-aluminum-vanadium alloy, JAAD Case Reports, 1 (2015) 36–37.

DOI: 10.1016/j.jdcr.2014.10.009

Google Scholar

[27] Devlin J.J., Pomerleau A.C., Brent J., Morgan B.W., Deitchman S., Schwartz M., Clinical features, testing, and management of patients with suspected prosthetic hip-associated cobalt toxicity: a systematic review of cases, Med Toxicol, 9 (2013).

DOI: 10.1007/s13181-013-0320-0

Google Scholar

[28] Miculescu F., Bojin D, Ciocan L.T., Antoniac I., Miculescu M, Miculescu N, Experimental researches on biomaterial-tissue interface interactions, JOAM, 9 (2007), 3303–3306.

DOI: 10.4028/www.scientific.net/kem.638.14

Google Scholar

[29] Niculescu M., Antoniac I., Blajan A., Metallic biomaterials processing technologies in order to obtain a new design for a hip prosthesis femoral component, Solid State Phenomena, 216 (2016) 239-242.

DOI: 10.4028/www.scientific.net/ssp.216.239

Google Scholar

[30] Nicoara M., Raduta A., Parthiban R., Locovei C., Eckert J., Stoica M., Low Young's modulus Ti-based porous bulk glassy alloy without cytotoxic elements, Acta Biomaterialia 36 (2016) 323–331.

DOI: 10.1016/j.actbio.2016.03.020

Google Scholar

[31] Ionescu R., Cristescu I., Dinu M., Saban R., Antoniac I., Vilcioiu D., Clinical, biomechanical and biomaterials approach in the case of fracture repair using different systems type plate-screw, KEM, 583 (2014) 150-154.

DOI: 10.4028/www.scientific.net/kem.583.150

Google Scholar

[32] Antoniac I., Sinescu C., Antoniac A., Adhesion aspects in biomaterials and medical devices, JAST, 30 (2016) 1-5.

Google Scholar

[33] Ionescu R., Mardare M., Dorobantu A., Vermesan S., Marinescu E., Saban R., Antoniac I., Ciocan D.N., Ceausu M., Correlation Between Materials, Design and Clinical Issues in the Case of Associated Use of Different Stainless Steels as Implant Materials, KEM, 583 (2014).

DOI: 10.4028/www.scientific.net/kem.583.41

Google Scholar

[34] Mao X., Wong A.A., Crawford R.W., Cobalt toxicity-an emerging clinical problem in patients with metal-on-metal hip prostheses?, Med J Aust, 194 (2011) 649-651.

DOI: 10.5694/j.1326-5377.2011.tb03151.x

Google Scholar

[35] Martin J.R., Camp C.L., Wyles C.C., Taunton M.J., Trousdale R.T., Lewallen D.G., Increased Femoral Head Offset is Associated With Elevated Metal Ions in Asymptomatic Patients With Metal-on-Polyethylene Total Hip Arthroplasty, J Arthroplasty, 31 (2016).

DOI: 10.1016/j.arth.2016.05.047

Google Scholar

[36] Caltana L., Merelli A., Lazarowski A., Brusco A., Neuronal and glial alterations due to focal cortical hypoxia induced by direct cobalt chloride (CoCl2) brain injection, Neurotox Res, 15 (2009) 348-358.

DOI: 10.1007/s12640-009-9038-9

Google Scholar

[37] Martin J.R., Spencer-Gardner L., Camp C.L., Stulak J.M., Sierra R.J., Cardiac cobaltism: a rare complication after bilateral metal-on-metal total hip arthroplasty-Case report, Arthroplasty Today, 1 (2015) 99-102.

DOI: 10.1016/j.artd.2015.10.002

Google Scholar

[38] Oldenburg M., Wegner R., Baur X., Severe Cobalt Intoxication Due to Prosthesis Wear in Repeated Total Hip Arthroplasty-Case Report, The Journal of Arthroplasty, 24 (2009) 15-20.

DOI: 10.1016/j.arth.2008.07.017

Google Scholar

[39] Pelclova D., Sklensky M., Janicek P., Lach K., Severe cobalt intoxication following hip replacement revision: Clinical features and outcome, Clin Toxicol, 50 (2012) 262-265.

DOI: 10.3109/15563650.2012.670244

Google Scholar

[40] Green B., Griffiths E., Almond S., Neuropsychiatric symptoms following metal-on-metal implant failure with cobalt and chromium toxicity, BMC Psychiatry, 17 (2017) 1-5.

DOI: 10.1186/s12888-016-1174-1

Google Scholar

[41] Machado C., Appelbe A., Wood R., Arthroprosthetic cobaltism and cardiomyopathy, Heart Lung Circ, 21 (2012) 759-760.

DOI: 10.1016/j.hlc.2012.03.013

Google Scholar

[42] Roessler P.P., Witt F., Efe T., Schmitt J., Arthroprosthetic cobaltism and pseudotumour also occur in patients with small diameter femoral ball head metal-on-metal total hip arthroplasties, BMJ Case Rep, 3 (2014) 1-4.

DOI: 10.1136/bcr-2013-203362

Google Scholar

[43] Tower S., Arthroprosthetic cobaltism: identification of the at-risk patient, Alaska Med, 52 (2010) 28–32.

Google Scholar

[44] Campbell P., Ebramzadeh E., Nelson S., Takamura K., De Smet K., Amstutz H.C., Histological features of pseudotumor-like tissues from metal-on-metal hips, Clinical Orthopedics and Related Research, 468 (2010) 2321-2327.

DOI: 10.1007/s11999-010-1372-y

Google Scholar

[45] Rotini R., Bettelli G., Cavaciocchi M., Savarino L., Pseudotumor due to metallosis after total elbow arthroplasty, Indian J Orthop, 51 (2017) 103-106.

DOI: 10.4103/0019-5413.197557

Google Scholar

[46] Romesburg J.W., Wasserman P.L., Schoppe C.H., Metallosis and Metal-Induced Synovitis Following Total Knee Arthroplasty: Review of Radiographic and CT Findings, J Radiol Case Rep, 4 (2010) 7-17.

DOI: 10.3941/jrcr.v4i9.423

Google Scholar

[47] Hsieh M.C., Yao M.S., Chan W.P., Metallosis, Intern Med, 51 (2012) 2071-(2072).

Google Scholar

[48] Goldenberg Y., Tee J.W., Salinas-La Rosa C.M., Murphy M., Spinal metallosis: a systematic review, Eur Spine J., 25 (2016) 1467-1473.

DOI: 10.1007/s00586-015-4347-6

Google Scholar

[49] Lukina E., Laka A., Kollerov M., Sampiev M., Mason P., Wagstaff P., Noordeen H., Yoon W.W., Blunn G., Metal concentrations in the blood and tissues after implantation of titanium growth guidance sliding instrumentation, Spine J., 16 (2016).

DOI: 10.1016/j.spinee.2015.11.040

Google Scholar

[50] Sayed-Noor A.S., Sjödén G.O., Severe metallosis after total elbow arthroplasty-a case report, Hand, 5 (2010) 86-89.

DOI: 10.1007/s11552-009-9201-z

Google Scholar

[51] Girard J., Soenen M., Monnin C., Migaud H., Bilateral simultaneous metal inlay dissociation from the polyethylene liner of a metal-on-metal hip replacement, Orthop Traumatol Surg Res., 95 (2009) 443-446.

DOI: 10.1016/j.otsr.2009.07.003

Google Scholar

[52] La Verde L., Fenga D., Spinelli M.S., Campo F.R., Florio M., Rosa M.A., Catastrophic metallosis after tumoral knee prosthesis failure: A case report, Int J Surg Case Rep, 30 (2017) 9-12.

DOI: 10.1016/j.ijscr.2016.11.023

Google Scholar

[53] Khan W.S., Agarwal M., Malik A.A., Cox A.G., Denton J., Holt E.M., Chromium, cobalt and titanium metallosis involving a Nottingham shoulder replacement, J Bone Joint Surg Br, 90 (2008) 502-505.

DOI: 10.1302/0301-620x.90b4.20302

Google Scholar

[54] Bradberry S.M., Wilkinson J.M., Ferner R.E., Systemic toxicity related to metal hip prostheses, Clin Toxicol, 52 (2014) 837-847.

DOI: 10.3109/15563650.2014.944977

Google Scholar

[55] Ene R., Panti Z., Nica M., Pleniceanu M., Ene P., Cirstoiu M., Antoniac I., Cirstoiu C., Mechanical failure of angle locking plates in distal comminuted tibial fractures, KEM, 695 (2016) 118-122.

DOI: 10.4028/www.scientific.net/kem.695.118

Google Scholar