Abstract:

Article Preview

In order to solve the problem of ECG acquisition and processing better, this paper designs a portable ECG tester, which uses the SCM as the core controller. Power basic frequency interference was removed by 50 Hz trap filter. Power harmonic interference, electrode polarization interference and my electricity interference were removed by low- pass filter. Baseline drift was removed by high- pass filter. At the same time; we test the performance of acquisition circuit, filter circuit and amplification circuit. As a result, it is shown that the system can well collect and process ECG signal.

Info:

Periodical:

Advanced Materials Research (Volumes 328-330)

Edited by:

Liangchi Zhang, Chunliang Zhang and Zichen Chen

Pages:

228-231

Citation:

X. H. Fu et al., "Design of Portable ECG Tester", Advanced Materials Research, Vols. 328-330, pp. 228-231, 2011

Online since:

September 2011

Export:

Price:

$38.00

[1] J. Li: The Design of A Wireless ECG Tele-Monitoring System(MS., BeiJing University of Tethnology, China 2008), p.12. (In Chinese).

[2] K.Y. Zhang: The Research on Remote ECG Monitoring System Basedon GPRS(MS., Harbin University of science and Tethnology, China 2008), p.8. (In Chinese).

[3] Information on http: /www. analog. com/static/imported-files/data_sheets/OP07. pdf.

[4] P. Wu: LCD Portable Myocardial Infraction Monitor(MS., Zhe Jiang University, China 2002), p.18. (In Chinese).

[5] H. Zh. Shi: The Research and Design of Wireless ECG Monitoring and Warning Device(MS., Ji Lin University, China 2007), p.21. (In Chinese).