Authors: Jung Hwa Hong, Sang Ok Ko, Soon Hyuck Lee
Abstract: "Osseo" refers to bone and "integration" refers to how a prosthesis can be integrated with
the bone in residual limbs both arms and legs. Osseointegration(OI) was originally defined as a
direct structural and functional connection between ordered living bone and the surface of a loadcarrying
implant. OI could be described as the modality for stable fixation of titanium implant to
bone structure. The OI has become a realized phenomenon of importance in the dental and
rehabilitation sciences since recently developed dentures and artificial limbs are directly attached to
human skeleton by using osseointegrated implants. Previously, a study showed that bone strain
generated potential (SGP) that is an electrical potential and considered to be generated by fluid flow
in bone could be used as a parameter to examine the amount of OI on bone-implant interface. Since
no study was performed to understand according to the point on behavior of SGP for the boneimplant
composite. In this study, we used a multi-channel measurement system and investigated
SGP according to the point. Four white New Zealand rabbits underwent pure titanium implant
insertion surgery to tibia after amputation. After checking full OI at the end of the 5 weeks,
experimental animals were euthanized and the amputated tibia-implants were harvested. Holes of
0.5mm in diameter were made on the tissue of the tibia and electrodes of 0.16mm in diameter were
inserted to the holes. Then, the instrumented bone-implant composites were placed to a servo
material testing machine to apply axial compressive displacement loadings. During the compression
tests, SGPs were also measured. Magnitude of SGP was found to be significantly increased near
bone and implant interface for the osseointegrated bone-implant composite.
1569
Authors: Sung Ki Min, Jung Hwa Hong, Sang Ok Ko
Abstract: Osseointegration (OI) could be described as the modality for stable fixation of titanium
implant to bone structure. The OI has become a realized phenomenon of importance in the dental
and rehabilitation sciences since recently developed dentures and artificial limbs are directly
attached to human skeleton by using osseointegrated implants. Previously, a study showed that bone
strain generated potential (SGP) that is an electrical potential and considered to be generated by
fluid flow in bone could be used as a parameter to examine the amount of OI on implant-bone
interface. SGP generation is known to require intraosseous fluid flow related with generations of
pore pressure gradient in bone. Therefore, SGP would interact with properties determining
interstitial fluid flow characteristics such as viscosity, velocity, flow path directions, and interstitial
fluid flow boundary conditions. Since interstitial fluid flow characteristics in bone are governed by
pore pressure gradient, it could be possible to predict SGP indirectly through the prediction of pore
pressure generation in bone. The aim of this study is to predict the distribution of pore pressure in
OI bone-implant composite representing a completely osseointegrated rabbit tibia-titanium implant
composite. The theoretical background of this prediction is based on the poroelasticity of 2-phase
material that grounds on fluid flow and behavior of cortical bone material. In this study, we
constructed a finite element (FE) model of the composite from images of micro-CT scanning. In the
next step, we examined analysis of the FE model about pore pressure by using ABAQUS. In this
analysis, the constitutive behavior was externally computed by utilizing a user subroutine. The
results showed the different spatial distributions of pore pressure in the composite. The magnitudes
of pore pressure were found to be significantly increased when the position was approached for the
interface of implant-bone. Further analytical study is required to fully understand relationships
between SGP and pore pressure distributions in OI bone-implant composite materials.
909
Authors: Jung Hwa Hong, Sang Ok Ko
Abstract: Osseointegration (OI) could be described as the modality for stable fixation of titanium
implant to bone structure. The OI has become a realized phenomenon of importance in the dental
and rehabilitation sciences since recently developed dentures and artificial limbs are directly
attached to human skeleton by using osseointegrated implants. Previously, a study showed that bone
strain generated potential (SGP) that is an electrical potential and considered to be generated by
fluid flow in bone could be used as a parameter to examine the amount of OI on implant-bone
interface. Since no study was performed to understand SGP behavior as a function of position for
the implant-bone composite, a one-dimensional map of SGP was constructed along the longitudinal
direction of the composite. For the purpose, nine electrodes including one reference were
instrumented on the wet composite for the one-dimensional mapping of SGP during compression
tests. The peak magnitudes of SGP were found to be significantly increased when the measurement
position was approached for the interface of implant-bone. The results could indicate that the spatial
SGP behavior of osseointegrated implant-bone composite could be caused by the interface of the
implant-bone.