Study on Calibration Technology for Helium Mass-Spectrometer Leak Detector

Article Preview

Abstract:

To calibrate the helium mass-spectrometer leak detector in-site in the whole range, a series of leaks with helium-permeable membranes of different leakage rates were connected to a helium mass-spectrometer leak detector separately, then a set of readings of standard leaks were gained, and the relation curves were fitted mathematically between the standard leakage rates and leakage readings so as to give a fitting formula between the whole-range readings and leakage rates. As a result, the simplicity and efficiency method make it suitable for calibrations of helium mass-spectrometer leak detector.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

360-363

Citation:

Online since:

December 2012

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2013 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] J.M. Hidalgo, J.L.de Segovia .Leak calibration by comparison with reference standard leaks[J]. Vacuum 82(2008),1151-1156.

DOI: 10.1016/j.vacuum.2008.01.050

Google Scholar

[2] Zhang Di-xin, Feng Yan,Zhang Jian-jun,Guo Mei-ru,Shen Hui. Calibration apparatus of comparison leak rate method[J].. Vcuum(in chinese) 2008,8, 14~17.

Google Scholar

[3] Charles D.Ehrich. Recommended practices for the calibration and use of leaks [J]. J.Vac.Sci.Technol .A10(1),Jan/Feb 1992. 1~17.

Google Scholar