Study on Calibration Technology for Helium Mass-Spectrometer Leak Detector
Abstract:
To calibrate the helium mass-spectrometer leak detector in-site in the whole range, a series of leaks with helium-permeable membranes of different leakage rates were connected to a helium mass-spectrometer leak detector separately, then a set of readings of standard leaks were gained, and the relation curves were fitted mathematically between the standard leakage rates and leakage readings so as to give a fitting formula between the whole-range readings and leakage rates. As a result, the simplicity and efficiency method make it suitable for calibrations of helium mass-spectrometer leak detector.
Info:
Periodical:
Pages:
360-363
Citation:
Online since:
December 2012