Detection of Phthalate Esters from Plastic Packaging Materials into Edible Oil by Gas Chromatography-Mass
Abstract:
A method for the simultaneous determination of 15 kinds of phthalate esters in edible oil by gas chromatography-mass spectrometer (GC-MS) was firstly established. The calibration curves showed good linearity in the range of 0.001~1mg /L and the correlation coefficients (r) were more than 0.999. The limits of quantification were in the range of 0.001~ 2.000 μg/L (S/N = 10). The recovery of three level ranged from 70.50% to 112.00%, at spiked levels of 0.1mg/L, 1.0mg/L and 10mg/L with the relative standard deviations (n = 6) of 0.59% ~ 7.54%. On the basis of the above method, the three edible oils under different storage temperatures were detected and the results showed that DIBP, DBP and DEHP were detected in two samples. This method possesses high sensitivity and low detection limit and can be used for quantitative analysis PAEs in edible oils.