Concept of the Partial Throw-Away Design of Piezoelectric Micropump for Medical Uses with Low Cost and Waste
Abstract:
In this study, piezoelectric micropumps with partial throw-away design have been developed and discussed. The actuator, a rectangular piezoelectric plate, can be separated from the pump body and be used repeatedly while the pump chamber is single-use to prevent infection. Thus, a low cost and waste solution for biomedical uses is achieved. A mechanical model is used to analyze the behavior of pump diaphragm and chamber geometry. Pump chamber with 20mm diameter is used due to its moderate elasticity. Then, 5 pumps with different depth of pump chamber have been made. The maximum flow rate, 5.7g/min, is reached by using a 0.5mm pump chamber with a 200V actuated voltage.
Info:
Periodical:
Pages:
363-367
Citation:
Online since:
December 2013
Authors:
Keywords:
Price:
Сopyright:
© 2014 Trans Tech Publications Ltd. All Rights Reserved