Biomechanical Investigation of the Retrograde Intramedullary Nail Fixation for Distal Femoral Periprosthetic Fracture after Total Knee Arthroplasty

Article Preview

Abstract:

This study investigated the effect of fixation screw number of intramedullary nail on the postoperative stability of the femoral construct in treatment of periprosthetic fractures after total knee arthroplasty. With use of a material testing machine, both compressive and torsion tests were conducted to evaluate the biomechanical behavior of intact, and femora with periprosthetic fractures treated with retrograde intramedullary nail with 5-screw and 9-screw fixations. The results indicated the followings: 1) the average compressive stiffness for intact, 5-screw and 9-screw groups were 1,159±61.7, 950.6±18.6 and 958.2±26.1 N/mm, respectively. No statistical difference was found between 5-screw and 9-screw groups (p>0.05). However, significant difference was found between intact/5-screw and intact/9-screw groups (p<0.05). 2). the average torsion stiffness for intact, 5-screw group and 9-screw group were 5,527±71.2, 2,652±35.4 and 2,858±81.7 N*mm/Degree, respectively. Statistical difference was found between each other groups (p<0.05).

You might also be interested in these eBooks

Info:

Periodical:

Pages:

389-394

Citation:

Online since:

October 2014

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2014 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

* - Corresponding Author

[1] F. Chen, M.A. Mont, R.S. Bachner, Management of ipisilateral supracondylar femur fracture following total knee arthroplasty. J Arthroplasty 9 (1994) 521-526.

DOI: 10.1016/0883-5403(94)90099-x

Google Scholar

[2] F.F. Jabczenski, M. Crawford, Retrograde intramedullary nailing of supracondylar femur fractures above total knee arthroplasty: a preliminary report of four cases. J Arthroplasty 10 (1995) 95-101.

DOI: 10.1016/s0883-5403(05)80108-1

Google Scholar

[3] R.W. Culp, R.G. Schmidt, G. Hanks, Supracondylar fracture of the femur following prosthetic knee arthroplasty. Clin Othorp. 222 (1987) 212-222.

DOI: 10.1097/00003086-198709000-00029

Google Scholar

[4] W.L. Healy, J.M. Siliski, S.J. Incavo, Operative treatment of distal femoral fractures proximal to total knee replacements. J Bone Joint Surg Am 75 (1993) 27-34.

DOI: 10.2106/00004623-199301000-00005

Google Scholar

[5] A. Cadambi, G.A. Engh, K.A. Dwyer, T.N. Vinh, Osteolysis of the distal femur after total knee arthroplasty. J Arthroplasty. 9 (1994) 579-594.

DOI: 10.1016/0883-5403(94)90111-2

Google Scholar

[6] D.D. Obertson, C.M. Mintzer, B.N. Weissman, Distal loss of femoral bone following total knee arthroplasty. J Bone Joint Surg Am 76 (1994) 66-76.

DOI: 10.2106/00004623-199401000-00009

Google Scholar

[7] M.C. Moran, G.W. Brick, C.B. Sledge, Supracondylar femoral fracture following total knee arthroplasty. Clin Orthop 324 (1996) 196-209.

DOI: 10.1097/00003086-199603000-00023

Google Scholar

[8] A.C. McLaren, J.A. Dupont, D.C. Schroeber, Open reduction internal fixation of supracondylar fractures above total knee arthroplasties using the intramedullary supracondylar rod. Clin Orthop 302 (1994) 194-198.

DOI: 10.1097/00003086-199405000-00030

Google Scholar

[9] J.A. Szivek, M. Thomas, J.B. Benjamin, Characterization of a synthetic foam as a model for human cancellous bone. J Appl Biomater 4 (1993) 269-272.

DOI: 10.1002/jab.770040309

Google Scholar

[10] A.D. Heiner and T.D. Brown, Structural properties of a new design of composite replicate femurs and tibias. J Biomech 34 (2001) 773-782.

DOI: 10.1016/s0021-9290(01)00015-x

Google Scholar

[11] J.A. Szivek and R.L. Gealer, Comparison of the deformation response of synthetic and cadaveric femora during simulated, one-legged stance. J Appl Biomater. 2 (1991) 277-280.

DOI: 10.1002/jab.770020410

Google Scholar

[12] M.M. Etersen, C. Olsen, J.B. Lauritzen, B. Lund, Changes in bone mineral density of the distal femur following uncemented total knee arthroplasty. J Arthroplasy 10 (1995) 7-11.

DOI: 10.1016/s0883-5403(06)80058-6

Google Scholar