Dose Rate Spatial Distribution Produced by the Pulsed X-Ray Source in the Radiographic Examination
Abstract:
In the article the dose rate measurement technique of the pulsed X-ray source RAP-160-5 is presented. The examples of the dose rate spatial distribution for the different pulsed X-ray tube parameters at the different distance between the pulsed X-ray source focus and the detector are demonstrated. The recommendations for the radiation dose optimization to the objects under investigation from the different distance between the pulsed X-ray source focus and the detector are proposed.