Hip Simulator Study of Ceramic-Ceramic and Ceramic-PE-Couplings with Different Head Diameters

Article Preview

Abstract:

The contact conditions and the resulting contact stresses for ceramic on ceramic pairings with diameter 28, 32 and 36mm were determined. A hip simulator according ISO 14242 was used to measure the wear behaviour of these hard – hard couplings in comparison with the common hard – soft pairing ceramic on polyethylene. It was found that there isn’t a significant dependence of the wear behaviour from articulation diameter for ceramic on ceramic but a strong increase of polyethylene wear with larger hip joint heads.

You might also be interested in these eBooks

Info:

Periodical:

Key Engineering Materials (Volumes 309-311)

Pages:

1269-1272

Citation:

Online since:

May 2006

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2006 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] Bader, R.; Willmann, G.: Keramische Pfannen für Hüftendoprothesen Teil 6: Pfannendesign, Inklinations- und Antetorsionswinkel beeinflussen Bewegungsumfang und Impingement. Biomed. Technik 44 (1999), 212-219.

DOI: 10.1515/bmte.1999.44.7-8.212

Google Scholar

[2] Bader, R.; Willmann, G.: Keramische Pfannen für Hüftendoprothesen Teil 7: Wie beeinflussen Lage des Rotationszentrums und CCD-Winkel des Schaftes Range of Motion und Impingement. Biomed. Technik 44 (1999), 345-351.

DOI: 10.1515/bmte.1999.44.12.345

Google Scholar

[3] Bader, R.; Scholz, R.; Steinhauser, E.; Mittelmeier, W.; Gradinger, R.: Verwendung großer Prothesenköpfe bei totalem Hüftgelenkersatz: Welche Vorteile ergeben sich für Bewegungsumfang und Luxationsstabilität?. Orthopäd. Praxis, (2002).

DOI: 10.1055/s-2003-821429

Google Scholar

[4] Oberbach, Th.; Salomon, D.; Glien, W.: Erhöhung des Range of Motion durch große Keramikköpfe und ein spezielles Design des Keramikinlays. DVM-Bericht 315; DVM 2004 ( 80); 37-48.

DOI: 10.3139/120.100653

Google Scholar

[5] ISO 6474: Implants for surgery - Ceramic materials based on high purity alumina, (1994).

Google Scholar

[6] ISO 14242 1/2: Implants for surgery - Wear of total hip joint prostheses.

Google Scholar

[7] Zagra, L.; Giacometti-Ceroni, R.; Crbella, M.: THA Ceramic-Ceramic Coupling: The Evaluation of the Dislocation Rate with Bigger Heads; in Lazennec, J. -Y.; Dietrich, M.: Bioceramics in Joint Arthroplasty - 9-th BIOLOX Symposium Proceedings, (2004).

DOI: 10.1007/978-3-7985-1968-8_28

Google Scholar

[8] Engelhardt, R.: Mehr Transparenz im Gesundheitswesen durch externe Qualitätssicherung?. Zeitschrift für Orthopädie und ihre Grenzgebiete; Bd. 142; (1/2004), 7-9 Medium wear rate.

Google Scholar