Design of a Bioelectrical Impedance Spectrometer Based on AD5933

Article Preview

Abstract:

This paper designed a bioelectrical impedance spectroscopy (BIS) measurement system based on the integrated impedance converter AD5933. The working principle of the AD5933 was briefly introduced, and a bipolar BIS measurement device was designed. The AD5933 can be controlled by MCU STC89LE516AD through an I2C bus. The software of the BIS device was written in C51 language and provides a friendly human-machine interface. Preliminary experiment showed that the impedance measurement error at different frequencies is less than 3% and found that the measurement accuracy is relatively high for larger resistor.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

392-396

Citation:

Online since:

December 2012

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2013 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] Passauer J, Petrov H, Leicht J, Schleser A and Pucalka K: 2009 Evaluation of clinical dry weight assessment in haemodialysis patients using bioimpedance spectroscopy: a cross-sectional study Nephrol Dial Transplant 16856.

DOI: 10.1093/ndt/gfp517

Google Scholar

[2] Whitman T A, Forrest J C, Morgan M T and Okos M R: 1996 Electrical measurement for detecting early postmortem changes in porcine muscle J. Anim Sci. Vol. 74, pp.80-90.

DOI: 10.2527/1996.74180x

Google Scholar

[3] Analog Devices Inc. 2005–2011 AD5933 Data Sheet. pp.1-44.

Google Scholar