Key Technique for Inner Hole Diameter Measure with Capacitive Probe

Article Preview

Abstract:

The capacitive probe for inner hole diameter measuring was introduced, the equipotential guard ring technique was adopted to eliminate edge effect of capacitive probe effective measuring electrode, the driving cables technique was used to reduce the influence of parasitic capacitance, and then we realize the precision measurement in any section of inner holes for diameter from Φ1.85 mm to Φ7 mm.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

827-830

Citation:

Online since:

February 2013

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2013 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] Hrioshi MURAKAMI, Akio KATSUKI, Hiromichi ONIKURA. Journal of Advanced Mechanical Design, Systems and Manufacturing, Vol. 4 (2010), pp.995-1004.

Google Scholar

[2] Stone J. Muralikrishnan B. Journal of Research of the National Institute of Standards and Technology, Vol. 116 (2011), pp.573-587.

Google Scholar

[3] Yu Yongxin, Zhang Heng, Wang Zongchao, Chang Yizhe. Transactions of Tianjin University, Vol. 16 (2010), pp.447-451.

Google Scholar

[4] Yu Yongxin, Zhang Heng, Chang Yizhe. Study of deep-hole aperture measurement management system based on capacitance sensor, 8th International Symposium on Test Measure, Vol. 1(2009), pp.1807-1811.

Google Scholar

[5] Wu Xiaofeng, Zhang Guoxiong. Nanotechnology and Precision Engineering, Vol. 4 (2006), pp.301-306.

Google Scholar

[6] Wang Baoguang, Zhang Heng, Qian Zhengyu. Acta Metrologica Sinica, Vol. 27(2006), pp.228-231, In Chinese.