Practice and Analysis of Life Measuring Device for Motor Valve

Article Preview

Abstract:

Motor valve leakage is not easy to detect during using, so it must carry out quality inspection before assembly into gas meter. The quality appraisement of valve mainly relies on switch times before leakage, and pressure difference measurement is better method for leakage judgment. The static pressure leak detection method, reference gas chamber and control program of solenoid valve are key technologies in pressure difference measurement and life measurement. The new type device is designed with hardware and software, such as gas path and pipes, sensor, the solenoid valves of dynamic structure static switching unit, controller, gas pressure regulation system, automatic detection and valve switching process control program etc. Based on support of hardware and software, dynamic flow pipe is separated to static segments which are reference gas chamber and measuring air chambers; and the signal of chambers pressure difference and switch times without leakage are the output result. Switching frequency ratio of solenoid valves to motor valve is 1:5 or 1:10; lower ratio can prolong the service life of solenoid valves.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

1972-1977

Citation:

Online since:

September 2013

Authors:

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2013 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] Beijing Institute of measuring and testing. JJG577-2005 Diaphragm gas meter testing procedures [S]. Beijing: Standards Press of China, (2005).

Google Scholar

[2] Ministry of Machinery Industry Institute of Shanghai Automation Instrumentation Industry. GBT6968-1997 Diaphragm Gas Meters [S]. Beijing: Standards Press of China, 1997.

Google Scholar

[3] QJ 1610-1989 Valve classification and detection of gas leakage rate [S]. Beijing: Standards Press of China, 1989.

Google Scholar

[4] DIN 3230-3-1982 Standard of Valve Leak Detection [S]. Berlin: Alleinverkaif der Normen durch Beuth Verlag GmbH, 1982.

Google Scholar

[5] R.A. Collacott, The identification of the source of machine noises contained within a multiple-source environment, Applied Acoustics, 1976, Volume 9(Issue 3): 225~238.

DOI: 10.1016/0003-682x(76)90020-7

Google Scholar

[6] Urban Construction Design Institute. CJ T112-2008 IC card diaphragm gas meter [S]. Beijing: Standards Press of China, (2008).

Google Scholar

[7] ASME B16. 104. Control Valve Seat Leakage [EB/OL]. http: /wenku. baidu. com/view/f5419030b90d6c85ec3ac657. html.

Google Scholar