Detection of L-Tert-Leucine in the Enzyme-Catalyzed Reaction System
Abstract:
Direct Determination of L-Ter-Leucine in Enzyme Catalytic Reaction System by HPLC Was Studied. the Detection Were Performed on a Kromasil 700-5C18 Column Using a Eluant Containing 0.25% (NH4)H2PO4 and 100% Methanol (V((NH4)H2PO4):V(methanol)=100:5) with the Flow Rate of 0.8 Ml/min at,detection Wavelength of 205nm. there Was a Good Line Correlation between Peak Area and Contents in the Rang of 0.2-10 Mg/ml, the Correlation Coefficient Was 0.9986, the Average Recovery Was 98.88% with a Relative Stand Deviation of 0.78% (n=5). this Method Is Simple, Stable, Accurate and Reliable for the Quality Control of L–ter-Leucine.