Investigation of Biomechanical and Biosafety of Injection Moulded Implant Materials

Article Preview

Abstract:

This paper presents the attempt to manufacture metallic implant using medical grade 316L stainless steel alloy powder by MIM process. The powder with the median particle size of 15 μm and a binder consisting of palm stearin and poly ethylene were mixed at 160°C using a sigma-blade mixer for one hour to prepare the feedstock of the test bar. The rheological properties of the feedstock was tested using capillary rheometer. The test bar was injection moulded using vertical injection moulding machine with the nozzle temperature of 200°C. Prior to sintering, the specimens were debound using a combination of solvent extraction and thermal pyrolysis method. The specimens were then sintered under vacuum at the temperature between 1300oC to 1360oC. The properties of the sintered bar such as physical appearance and densities were presented and discussed. The biocompatibility including toxicity properties of the implant also been presented. The results showed that physical and mechanical properties of the sintered sample complied with the international standard

You might also be interested in these eBooks

Info:

Periodical:

Pages:

79-84

Citation:

Online since:

January 2014

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2014 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] R.M. Germa and A. Bose, Injection Moulding of Metals and Ceramics, MPIF, Princeton, NJ, 1997.

Google Scholar

[2] R.M. German and R. Cornwall, The Powder Injection Moulding Industry: An Industry and Market Report. Innovative Materials Solution Inc, Pennsylvania, 1997.

Google Scholar

[3] R.M. German, K.F. Hens and S.T. Lin, Ceramic Bulletin. Vol.70 ( 8), 1294-1301, 1991.

Google Scholar

[4] M.A. Omar, H.A. Davies, P.F. Messer and B.Ellis, Journal of Materials Processing Technology 113: 477-481, (2001)

Google Scholar

[5] Standard Test Methods for Metal Powder and Powder Metallurgy Products, MPIF, Edition, Princeton, NJ , (1992)

Google Scholar

[6] ISO/EN 10993-5: biological evaluation of medical devices – part 5 : Test for Cyctotoxicity: in-vitro Method

Google Scholar

[7] Zulkifly , A.H. Test For In Vivo Study 316l Stainless Steel Metal Injection Molding Technique - Animal Surgery Evaluation (Grossing Observation), (Report BT-02-01-12 –Techno Fund 1208D168), 2012.

Google Scholar