Clinical Results for 19 to 21 Years on Total Hip Prosthesis with Ceramic Head Combined with UHMWPE Socket

Article Preview

Abstract:

In total hip prosthesis (THP), we have been using alumina ceramic femoral heads to reduce polyethylene debris and the modified cementing technique, which was named “interface bioactive bone cement (IBBC)” method to improve implant fixation. In this study, we investigated the long-term clinical performance of THPs with an alumina ceramic head and IBBC method. From 1986 to 1988, 285 joints (215 patients) were operated on by a senior surgeon, and 265 joints (192 patients) could be followed up. The presence of radiolucent lines, loosening, osteolysis and ultrahigh-molecular-weight polyethylene (UHMWPE) socket wear were observed. In IBBC, a radiolucent line appeared as a ’space’, and loosening appeared as a ’separation’ between the HA layer and the cement. A ‘space’ appeared in 3 joints (1.4%) on the acetabulum and in 4 joints (1.8%) on the femur, while a ‘separation’ appeared in 3 joints (1.4%) on the acetabulum. Osteolysis was noted in 1 joint (0.5%) on the acetabulum and in 2 joints (0.9%) on the femur. No revision surgery was required. In our previous study, we reported that the thickness of the socket affected its clinical wear rate and that the wear rate of sockets with an alumina ceramic head was 20% lower than that of sockets with a metal head. By reducing wear debris through the use of ceramic heads, osteolysis could be reduced. Long-term fixation of THPs to the bone has been achieved by using IBBC. Thus, the long-term clinical results of THPs with an alumina head that were fixed using IBBC were excellent.

You might also be interested in these eBooks

Info:

Periodical:

Key Engineering Materials (Volumes 361-363)

Pages:

1303-1306

Citation:

Online since:

November 2007

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2008 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] Oonishi H, Wakitani S, Murata N, et al. Clinical Orthopaedic and Related Reserch, 379 (2000), 77-84.

Google Scholar

[2] Oonishi H, Kadoya Y, Iwaki H, et al. J of Arthroplasty, 16(2000), 784-789.

Google Scholar

[3] Oonishi H, Tsuji E and Kim YY, J Mat. Sci Mat in Med (1998) 393- 401.

Google Scholar