Comparison Measurement between Fizeau Interferometer and Scanning Deflectomtric Profiler

Article Preview

Abstract:

There is a strong demand to measure an absolute surface profile with a few nanometer uncertainty. The National Metrology Institute of Japan (NMIJ) measure the absolute surface profile using two instruments based on difference methods: the Fizeau type flatness interferometer and the scanning deflectometric profiler. In the Fizeau type flatness interferometer, the measurand is a gap distance between a reference flat and a specimen. To obtain an absolute profile of the specimen, the reference flat was evaluated by a three flat test and a FEM analysis. The three flat test is one of the methods to measure absolute profile of a reference flat; however, the reference flat deforms under the force of gravity, and its absolute value cannot be detected by the three flat test. The deformation value of the reference flat was calculated by the FEM analysis. In the scanning deflectometric profiler, a local slope angle of a specimen is measured using an autocollimator and the surface profile is obtained by integrating with the local slope angles. To obtain an absolute profile of the specimen, the autocollimator which is key elements for the scanning deflectometric profiler was evaluated by using the self-calibration type rotary table. In this paper, we performed the comparison measurement between the Fizeau type flatness interferometer and the scanning deflectometric profile and verified the accuracy of the two instruments based on the different methods.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

310-315

Citation:

Online since:

August 2014

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2015 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

* - Corresponding Author

[1] G. Schulz G and J. Schwider, Precise Measurement of Planeness, Applied Optics, 6 (1967) 1077-1084.

DOI: 10.1364/ao.6.001077

Google Scholar

[2] T. Takatsuji, S. Osawa, Y. Kuriyama and T. Kurosawa, Stability of the reference flat used in Fizeau interferometer and its contribution to measurement uncertainty, SPIE, 5190 (2003) 431-439.

DOI: 10.1117/12.505537

Google Scholar

[3] R. D. Geckeler, Optimal use of pentaprisms in highly accurate deflectometric scanning, Meas. Sci. Technl., 18 (2007) 115-125.

DOI: 10.1088/0957-0233/18/1/014

Google Scholar

[4] T. Watanabe and H. Fujimoto, High accuracy angle index table, XVIII IMEKO (2006).

Google Scholar