Evaluation of Measurement Uncertainty for Polymer Thin Film Humidity Sensor Using CMH and PRT Based on Dew Point Temperature Variation

Article Preview

Abstract:

The purpose of this paper is to propose appropriate evaluation method of measurement uncertainty with respect to the measurement results of polymer thin film capacitive humidity sensor according to dew point temperature variation. All of the evaluation process concerning the measurement uncertainty of polymer thin film capacitive humidity sensor is based on experiment and numerical analysis. To estimate uncertainty contribution properly for polymer thin film capacitive sensor strongly influenced by measurement environment, the experiments are performed in the reference relative humidity range from 5 % to 30 % at a setting dry bulb temperature of 23°C. Also, the reference standards are chilled mirror hygrometer (CMH) and platinum resistance thermometer (PRT). In particular, the uncertainty sources are considered differently according to the range of below or above 0°C, reference dew point. From the budgets, uncertainty contributions to the standard uncertainty are clearly dominated by the effect of the uncertainty due to super-cooling in case of condition below 0°C, dew point temperature. Also, the contributions for stability and distribution of humidity generator have a mainly effect on the expanded uncertainty of polymeric humidity sensor. It is desired that measurement uncertainties of polymeric humidity sensor are comprised separately on the basis of reference dew point, 0°C. The evaluation procedure of measurement uncertainty is developed in accordance with the principles established in the documents EA-4/02, Expression of the Uncertainty of Measurement in Calibration [1].

You might also be interested in these eBooks

Info:

Periodical:

Pages:

587-594

Citation:

Online since:

February 2017

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2017 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

* - Corresponding Author

[1] Expression of the Uncertainty of Measurement in Calibration, EA-4/02, (1999).

Google Scholar

[2] T. H. Lu, C. C. Chen, Uncertainty evaluation of humidity sensors calibrated by saturated salt solutions, Measure. 40 (2007) 591–599.

DOI: 10.1016/j.measurement.2006.09.012

Google Scholar

[3] E. EL-Din, M. Mahmoud, Realization of humidity standard facility using two-pressure humidity generator and humidity chamber, Metrol. 2009, Index 330930.

Google Scholar

[4] B. Cretinon, The CETIAT Standard Humidity Generator Operating in dew point temperature, Third International Symposium on Humidity & Moisture, UK, (1998).

Google Scholar

[5] H. Venzke, A. Melling, Papers and Abstracts from Third International Symposium on Humidity & Moisture, 1 (1998) 476-477.

Google Scholar

[6] H. G. Liedberg, M. R. Mnguni, D. Jonker, A Simple Humidity Generator for Relative Humidity Calibrations, Int. J. Thermophys. 29 (2008) 1660–1667.

DOI: 10.1007/s10765-008-0423-z

Google Scholar

[7] J. Lovell-Smith, The propagation of uncertainty for humidity calculations, Metrol. 46 (2009) 607–615.

Google Scholar

[8] Standard Test Method of Measuring Humidity with Cooled-Surface Condensation Hygrometer, ASTM, D4230-02 (Reapproved 2012).

Google Scholar

[9] Standard Test Methods for Water Vaper Content of Electrical Insulating Gases by Measurement of Dew Point, ASTM, D2029-97, (2008).

Google Scholar

[10] M. Heinonen, M. Vilbaste, Frost-point Measurement Error Due to a leak in a Sampling Line, Int. J. Thermophys. 29 (2008) 1589–1597.

DOI: 10.1007/s10765-007-0345-1

Google Scholar

[11] D. Zvizdic, M. Heinonen, New primary low-range dew point generator at LPM, IMEKO World congress, Lisbon, Portugal, 2009, p.1567–1570.

Google Scholar

[12] Gas analysis-Determination of the water dew point of natural gas-Cooled surface condensation hygrometers, ISO 6327.

DOI: 10.3403/00170683

Google Scholar