Investigations of In Vitro Corrosion, and Wear Properties of TiN PVD Coating on Ti6Al4V Alloy for Dental Application

Article Preview

Abstract:

Ti alloys have been widely used for biomedical applications.The generated wear debris due to its low wear resistance lead to inflammations and reduce the lifetime of the implant. So, the coating has been used to improve corrosion and wear resistance. In the present work, we report the assessments of TiN coating deposited by CA-PVD on Ti6Al4V alloy for dental applications. The deposited coating was characterized by XRD and SEM. Mechanical properties were assessed using micorindentation. The in vitro corrosion properties was investigated using PDP and EIS in artificial saliva. The wear rate was characterized using the linear reciprocating tribometer. The results showed that a TiN layer with 1.8 µm thickness was deposited with a hardness of 24.9±8.123 GPa and a Modulus of elasticity of 244.7±14.8 GPa. The corrosion study revealed enhancement in the corrosion performance of the coated sample in artificial saliva. The wear rate of the coated samples was also enhanced for the coated sample. The improved surface hardness, corrosion and wear characteristics of the TiN coating suggests that TiN coating would be a potential candidate for dental applications.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

1-6

Citation:

Online since:

July 2019

Keywords:

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2019 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

* - Corresponding Author

[1] Mohamed A. Hussein, Madhan Kumar, Robin Drew, Nasser Al-Aqeeli, Electrochemical Corrosion and In vitro Bioactivity of Nano-Grained Biomedical Ti-20Nb-13Zr Alloy in a Simulated Body Fluid, materials,11 (2017) 26.

DOI: 10.3390/ma11010026

Google Scholar

[2] Kamachimudali, U., T. M. Sridhar, and Baldev Raj. , Corrosion of bio implants., Sadhana 28, no. 3-4 (2003) 601-637.

DOI: 10.1007/bf02706450

Google Scholar

[3] M. A. Hussein, M. Azeem, A. Madhan Kumar, N. Al-Aqeeli,N. K. Ankah, A.A. Sorour, Influence of thermal treatment on the microstructure, mechanical properties, and corrosion resistance of newly developed Ti20Nb13Zr biomedical alloy in a simulated body environment, Journal of Materials Engineering and Performance,28( 2019) 1337–1349.

DOI: 10.1007/s11665-019-03908-4

Google Scholar

[4] Licausi, MARIE-PIERRE, A. Igual Muñoz, and V. Amigó Borrás. Influence of the fabrication process and fluoride content on the tribocorrosion behavior of Ti6Al4V biomedical alloy in artificial saliva., Journal of the mechanical behavior of biomedical materials 20 (2013) 137-148.

DOI: 10.1016/j.jmbbm.2013.01.019

Google Scholar

[5] N. Schiff, B. Grosgogeat, M. Lissac, and F. Dalard, Influence of Fluoride Content and pH on the Corrosion Resistance of Titanium and Its Alloys, Biomaterials, 23(9)(2002) 1995–(2002).

DOI: 10.1016/s0142-9612(01)00328-3

Google Scholar

[6] Chandra, A., J. J. Ryu, P. Karra, P. Shrotriya, Viggo Tvergaard, M. Gaisser, and T. Weik. Life expectancy of modular Ti6Al4V hip implants: influence of stress and environment., Journal of the mechanical behavior of biomedical materials 4(8) (2011) 1990-2001.

DOI: 10.1016/j.jmbbm.2011.06.018

Google Scholar

[7] Zheng, Yufeng, Dong Liu, Xiliang Liu, and Li Li. Ti–TiC–TiC/DLC gradient nano-composite film on a biomedical NiTi alloy., Biomedical Materials 3(4) (2008) 044103.

DOI: 10.1088/1748-6041/3/4/044103

Google Scholar

[8] M. Geetha, U. Kamachi Mudali, N. D. Pandey, R. Asokamani & Baldev Raj, Microstructural and Corrosion Evaluation of Laser Surface Nitrided Ti-13Nb-13Zr Alloy, Surf. Eng, 20(1)(2004) 68-74.

DOI: 10.1179/026708404225010595

Google Scholar

[9] Xiaomei Wang, and Hau-Chung Man., Enhancing the cell proliferation performance of NiTi substrate by laser diffusion nitriding., Surf. Coat. Technol, 309(2017) 59-66.

Google Scholar

[10] S. Shimadaa, T. Y. Takadaa, and J. Tsujino,Deposition of TiN films on various substrates from alkoxide solution by plasma-enhanced CVD, Surf. Coat. Technol., 199(1)(2005) 72–76.

DOI: 10.1016/j.surfcoat.2005.02.092

Google Scholar

[11] Adesina, Akeem Yusuf, Zuhair M. Gasem, and Arumugam Madhan Kumar. Corrosion resistance behavior of single-layer cathodic arc PVD nitride-base coatings in 1M HCl and 3.5 pct NaCl Solutions., Metallurgical and Materials Transactions B 48 (2017) 1321-1332.

DOI: 10.1007/s11663-016-0891-7

Google Scholar

[12] W.C. Oliver and G.M. Pharr, An Improved Technique for Determining Hardness and Elastic Modulus Using Load and Displacement Sensing Indentation Experiments, J. Mater. Res., 7(6)(1992) 1564–1583.

DOI: 10.1557/jmr.1992.1564

Google Scholar

[13] Hussein, Mohamed A., Bekir Yilbas, A. Madhan Kumar, Robin Drew, and Nasser Al-Aqeeli. Influence of Laser Nitriding on the Surface and Corrosion Properties of Ti-20Nb-13Zr Alloy in Artificial Saliva for Dental Applications., Journal of Materials Engineering and Performance 27, no. 9 (2018) 4655-4664.

DOI: 10.1007/s11665-018-3569-2

Google Scholar

[14] Ying Li, Li Qu, Fuihui Wang,The electrochemical corrosion behavior of TiN and (Ti,Al)N coatings in acid and salt solution,Corrosion Science,Volume 45, Issue 7,2003,Pages 1367-1381.

DOI: 10.1016/s0010-938x(02)00223-8

Google Scholar

[15] M.A. Hussein, A.S. Mohammed, N. Al-Aqeeli, Wear characteristics of metallic biomaterials: a review,Materials 8 (2015) 2749–276.

DOI: 10.3390/ma8052749

Google Scholar