In Vivo Failure Phenomena of Ultra-High Molecular Weight Polyethylene Acetabular Cups

Article Preview

Abstract:

During this study the distortion phenomenon was investigated on selected UHMWPE cups removed from patients, using 3dimensional measurement method. The cups investigated were machined by Metrimed Ltd. from ram extruded Chirulen 1020 type UHMWPE rods. The samples were measured with a 3D touch coordinate measurement device with an accuracy of 3 microns.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

521-526

Citation:

Online since:

November 2012

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2013 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] Kurtz S.M. The UHMWPE Handbook, Elsevier Academic Press, London, (2004).

Google Scholar

[2] Li S., Burstein AH. Ultra-high molecular weight polyethylene. The material and its use in total joint implants. J Bone Joint Surg Am; 76 (1994) 1080-90.

DOI: 10.2106/00004623-199407000-00020

Google Scholar

[3] ISO 14242: 2002. Implants for surgery – Wear of total hip-joint prostheses – Part1: Loading and displacement parameters for wear-testing machines and corresponding environmental conditions for test.

DOI: 10.3403/02535243u

Google Scholar

[4] ISO 14242: 2002. Implants for surgery – Wear of total hip-joint prostheses – Part2: Methods of measurement.

Google Scholar

[5] ASTM F 1714 – 96: 2002. Standard Guide for Gravimetric Wear Assessment of Prosthetic Hip-Designs in Simulator Devices.

DOI: 10.1520/f1714-96r18

Google Scholar

[6] Teeter, M.G.; Naudie, D.D. R; Carron, K.D.; Holdsworth, D. W.,. Three-dimensional surface deviation maps for analysis of retrieved polyethylene acetabular liners using micro-computed tomography. The J of Arthro 25 (2010) 330-332.

DOI: 10.1016/j.arth.2009.11.001

Google Scholar

[7] Wu, J.S.; Hsu, S.; Chen, J.,. Wear patterns of, and wear volume for, hemispherical acetabular cup liners. Wear 268 (2010) 481-487.

DOI: 10.1016/j.wear.2009.09.007

Google Scholar

[8] Champion AR., Li S., Saum K., Howard E., Simmons W. The effect of crystallinity on the physical properties of UHMWPE. Trans. of the 40th Orth Res Soc 19 (1994) 585.

Google Scholar

[9] Oonishi H. Long term clinical results of THR. Clinical results of THR of an alumina head with a cross-linked UHMWPE cup. Orth Surg and Traum 1995; 38: 1255-1264.

Google Scholar

[10] Martin, P.E., Nelson, R.C. The effect of carried loads on the walking pattern of men and women. Ergon 29 (1986) 1191-1202.

DOI: 10.1080/00140138608967234

Google Scholar

[11] Hurd, W.J.; Chmielewski, T.L.; Axe, M.J. Davis, I.; Snyder-Mackler, L.: Differences in normal and perturbed walking kinematics between male and female athletes. Clin Biomech 19 (2004) 465-472.

DOI: 10.1016/j.clinbiomech.2004.01.013

Google Scholar

[12] http: /www. orthopaedics. com. sg/treatments/total-hip-replacement.

Google Scholar