Zero Error Analysis and Compensation of No-Beat Time Division Multiplier

Article Preview

Abstract:

Time division multiplier was widely used in high-precision industrial frequency AC power measurements.The poor stability of zero point was caused by the switch leakage and amplifier offset voltage. Based on the experimental study of the time division multiplier circuit designed in the article, the mathematical model of zero point compensation was established, and then the appropriate calibration method was proposed. Considered the amplifier offset voltage, the error model of the time division multiplier was analyzed and the error expression was obtained. These can provide theoretical guidance for choosing amplifiers of the no-beat time division multiplier.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

3293-3297

Citation:

Online since:

October 2011

Authors:

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2012 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] Piotrs. Filipski. Watt and Voltampere Transfer Standard. IEEE Trans[M]. LM VOL 42. NO. 2. April (1993).

Google Scholar

[2] Guo Yongkun. The principle of Time division multiplier. Electrical Measurement & Instrumentation. (04): 1-7, (1978).

Google Scholar

[3] Wang Xuewei, Jiang Zhen, Su Quanzhi. The evaluation of errors about time division multiplier with hysteresis comparator in AC power measurement[J]. Electrical Measurement& Instrumentation. 37(414): 5-7, (2000).

Google Scholar

[4] Na Shuzhi, Liu Dexin. The principle of Time division multiplier for measuring the power and design and calculation of practical circuit[J]. Electrical Measurement &Instrumentation. (11):13-18, (1989).

Google Scholar