p.954
p.958
p.962
p.968
p.972
p.978
p.983
p.987
p.992
Research on the Pressure-Measuring Uncertainty of Standard Copper-Cylinder
Abstract:
This article introduces the method of selection and related criterion of standard copper-cylinder, and systematically analyzes five factors that affect the measuring uncertainty of standard copper-cylinder: primary pressure standard, height measuring of copper-cylinder, and inconsistent individual deformation of standard copper-cylinder, inconsistent pressure sensitivity and the influence of temperature. According to usage characteristics and selection methods of the standard copper-cylinder, discusses computing method of components of the measurement uncertainty and models measuring uncertainty of standard copper-cylinder. The model can quantitatively calculate through experimental data of selection, which lay a theoretical foundation for the control of the pressure measuring uncertainty of standard copper-cylinder.
Info:
Periodical:
Pages:
972-977
Citation:
Online since:
December 2012
Authors:
Price:
Сopyright:
© 2013 Trans Tech Publications Ltd. All Rights Reserved