Design of Labview Based Equipment for Measuring Thermally Stimulated Current

Article Preview

Abstract:

In this paper, we improved the home-made EL experimental set-up, have increased the function of TSC, enable the joint measurement of EL and TSC to measure the trap of shallow and deep. It is significant to study material properties. through the Labview8.6 software programming, the data of temperature and current could be collected, then we can get the curve that the current changes with temperature, it is the TSC curve. Experiment result proved the reliability of laboratory equipment.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

555-558