Effect of Continuous Arterial Blood Flow of Intra-Aorta Pump on the Aorta - A Computational Study

Article Preview

Abstract:

In this study, a computational fluid dynamics (CFD) study based on a finite element method (FEM) was performed for the human aorta with four different flow time patterns (healthy to full intra-aorta pump support). Fully coupled fluid-solid interaction (FSI) simulation was used to investigate the flow profiles in the aortic arch and its branches where the maximum disturbed and non-uniform flow patterns, and the wall shear stress profiles on the same areas. The blood flow was assumed as a homogeneous, incompressible, and Newtonian fluid flow. Flow across four inlets of aortas was derived from a lumped parameter model (LPM). The inlet flow rate waveforms were divided by different blood assist index (BAI), and were calculated with the physiological information of a heart failure patient.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

672-676

Citation:

Online since:

January 2013

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2013 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] Y. Chang, B. Gao, "A Global Sliding Mode Controller Design for an Intra-Aorta Pump, " ASAIO JOURNAL,USA,56(6),pp:510-516, December 2010.

DOI: 10.1097/mat.0b013e3181ede369

Google Scholar

[2] B Gao, KY Gu, Y Zeng, et al, "An Anti-Suction Control for an Intra-Aorta Pump Using Blood Assistant Index : A Numerical Simulation," ARTIFICIAL ORGANS, 36(3), PP:275-U183, May 2012.

DOI: 10.1111/j.1525-1594.2011.01309.x

Google Scholar

[3] B Gao, KY Gu, Y Zeng, et al, "A Blood Assist Index Contral by Intraaorta Pump : A Control Strategy for Ventruculat Recovery," ASAIO JOURNAL, USA, 57(5),pp:358-462, October 2011.

DOI: 10.1097/mat.0b013e3182257fac

Google Scholar

[4] YJ Xuan, KY Gu, Y Chang, et al, "A Contactless Power Supply System for Intra-corporeal Medical Instrumentations," 2011 International Conference on Multimedia and Signal Processing (CMSP), pp:3-6,2011.

DOI: 10.1109/cmsp.2011.9

Google Scholar

[5] YJ Xuan, Y Chang, KY Gu, et al, "Hemodynamic Simulation Study of a Novel Intra-Aorta Left Ventricular Assist Decice," ASAIO JOURNAL, USA, 58(5), pp: 462-469, September, 2012.

DOI: 10.1097/mat.0b013e318268eaf7

Google Scholar

[6] B Gao, Y Chang, KY Gu, et al, "A Pulsatile Control Algorithm of Continious-Flow Pump For Heart Recovery," ASAIO JOURNAL, USA, 58(4),pp :343-352, August 2012.

DOI: 10.1097/mat.0b013e318256bb76

Google Scholar

[7] Y.C. Fung, "Biomechanics : Circulation," Springer-Verlag, New York, 1996.

Google Scholar

[8] Li.John K-J, "Dynamics of the Vascular System," Word Scientific Publishing Co, Singapore,2004.

Google Scholar

[9] O.Hanieh Niroomand, T.S. Mohammad,G. Farzan,"Biomechanical analysis of wall remodeling in elastic arteries with application of fluid-solid interaction methods,'' JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY,7(4),pp:433-447, December 2007.

DOI: 10.1142/s0219519407002418

Google Scholar

[10] B.B Daily, T.W Pettitt, S.P. Sutera, W.S. Pierce, "Pierce-Donachy Pediatric VAD: Progress in Development,'' Annals of Thoracic Surgery, 61(1),pp:437–443, January 1996.

DOI: 10.1016/0003-4975(95)00991-4

Google Scholar

[11] P. Hochareon, K.B. Manning, A.A. Fontaine J.M. Tarbell, S. Deutsch, "Fluid Dynamic Analysis of the 50 CC Penn State Artificial Heart Under Physiological Operating Conditions Using Particle Image Velocimetry,'' American Society of Mechanical Engineers,USA 126(5), pp : 585–593, October 2004.

DOI: 10.1115/1.1798056

Google Scholar

[12] A. Untaroiu, H.G. Wood, P.E. Allaire, et al, "Computational Design and Experimental Testing of a Novel Axial Flow LVAD,'' American Society for Artificial Internal Organs, USA, 51(6),pp :702-710,December 2005.

DOI: 10.1097/01.mat.0000186126.21106.27

Google Scholar

[13] S. Deutsch, J.M. Tarbell, K.B. Manning, G. Rosenberg, A.A. Fontaine, "Experimental Fluid Mechanics of Pulsatile Artificial Blood Pumps,'' ANNUAL REVIEW OF FLUID MECHANICS, USA, 38,pp:65–86, 2006.

DOI: 10.1146/annurev.fluid.38.050304.092022

Google Scholar

[14] K.B. Manning, B.D. Wivholm, N. Yang, A.A. Fontaine, S. Deutsch, "Flow Behavior Within the 12-CC Penn State Pulsatile Pediatric Ventricular Assist Device: An Experimental Study of the Initial Design,'' Artificial Organs, USA, 32(6),pp:442–452, June 2008.

DOI: 10.1111/j.1525-1594.2008.00565.x

Google Scholar

[15] S.J. Sherwin, O. Shah, D.J. Doorly, S.J. Sherwin, "The Influence of Out-of-Plane Geometry on the Flow Within a Distal End-to-Side Anastomosis,'' Journal of Biomechanical Engineering-transactions of the ASME, England, 35(9),pp:86–95, September 2002.

DOI: 10.1115/1.429630

Google Scholar

[16] A.J.C. Ladd, "Numerical simulations of particulate suspensions via a discretized Boltzmann equation. Part 1. Theoretical foundation,'' Journal of Fluid Mechanics, 271, pp:209–285, 1994.

DOI: 10.1017/s0022112094001771

Google Scholar

[17] A.J.C. Ladd, "Numerical simulations of particulate suspensions via a discretized Boltzmann equation. Part 2. Numerical results,'' Journal of Fluid Mechanics, 271,pp:286–339, 1994.

DOI: 10.1017/s0022112094001783

Google Scholar

[18] D.R. Noble, J.R. Torczynski, "A lattice Boltzmann method for partially saturated computational cells,'' International Journal of Modern Physics C, Singapore, 9(8),pp:1189–1201, December 1998.

DOI: 10.1142/s0129183198001084

Google Scholar

[19] Y.C. Qiu, J.M. Tarbell, "Numerical simulation of pulsatile flow in a compliant curved tube model of a coronary artery,'' . J Biomech Eng , USA ,122(1),pp: 77–85, Feburary 2000.

DOI: 10.1115/1.429629

Google Scholar

[20] H.F. Younis, M.R. Kaazempur-Mofrad, R.C. Chan, et al, "Hemodynamics and wall mechanics in the human carotid bifurcation and its consequences for atherogenesis: Investigation of inter-individual variation,'' Biomech Model Mechanobiology, USA, 3(1),pp: 17–32, April, 2004.

DOI: 10.1007/s10237-004-0046-7

Google Scholar

[21] Y.C. Fung, "Biomechanics: Circulation, 2nd ed,'' Springer-Verlag New York, 1996.

Google Scholar

[22] Y.K. Sui, B.S. Ma, B. Chu, Q. A.K. Qiao, "Optimization of Anastomotic Configuration in CABG Surgery,'' Communications in Numerical Methods in Engineering, England, 25(11),pp :1097-1106, November 2009.

DOI: 10.1002/cnm.1310

Google Scholar