Background Leakage of Pressure Leak Calibration Caused by Temperature Change
Abstract:
On laboratory conditions,background leakage caused by temperature change was one of the most important factors for affecting calibration of pressure leak, consequently, the constant temperature water bath of the calibration apparatus has been used to reduce background leakage by means of keeping the apparatus temperature constant. The method of active and passiveness were combined to keep constant temperature. through the research on experimentation, it was found that background leakage was decreased to the grade of 1.36×10-9Pam3/s when constant temperature been used, and meanwhile the lower limiting of pressure leak was extended to 2.667×10-8Pam3/s.
Info:
Periodical:
Pages:
339-343
Citation:
Online since:
February 2013