Circuit Design for Alternating Current Field Measurement Make Use of Lock-in Amplifier

Article Preview

Abstract:

In ACFM, the measured signal is very weak and always inevitably includes the noises and the unwanted signal. Make use of lock-in amplifier technology to have designed a disturbance magnetic field testing circuit which includes eliminating 2.5 volt DC circuit, amplifying and filtering circuit, phase-shifting circuit and switching voltage multiplier. This circuits designed can performe the measurement and do-nosing for weak magnetic signal of ACFM. The experiment indicated that make use of lock-in amplifier technology to design and optimize ACFM detection circuit is an effective method