Flow Injection On-Line Distillation Method for Determination of Cyanide in Water
Abstract:
Purpose The flow injection distillation (zinc acetate solution) was changed to be 5% phosphoric acid in the method of determination of cyanide. Application Cyanide in groundwater was measured by FIA6000 + automatic flow injection analyzer directly. Results In the 2.0-100.0μg/L range, the linear correlation coefficient was greater than 0.9, the recovery was 86.1% -100.8%, relative standard deviation 1.3% -3.8%. Conclusion This method not only reduces the white crystals [Zn (OH)2] on the pump tube blockage, making the operation more simple and efficient, but also for the detection of a large number of groundwater samples.