A Method for In Situ Testing and Calibration of Moving-Coil Geophones
Abstract:
A new method for measuring and identifying characteristic parameters of moving-coil geophones is proposed in this paper, where sine sweeping signal is considered as the electric excitation instead of the traditional vibrational excitation. A mathematical model of moving-coil geophone is constructed and the frequency response of 6-parameter is obtained in this paper. Using the expression as the objective function for fitting measured amplitude-frequency curve, geophones characteristic parameters are obtained. Finally, verification test is performed and the factors affecting measure accuracy are analyzed. The experimental results indicate that the presented method is suitable for production and maintenance, especially for in-situ calibration of geophones. This work provides a new way to design and manufacture portable geophone parameter testers.
Info:
Periodical:
Pages:
1188-1192
Citation:
Online since:
August 2013
Authors:
Price:
Сopyright:
© 2013 Trans Tech Publications Ltd. All Rights Reserved
Share:
Citation: