Preparation of Sorbent Tube Certified Reference Materials

Article Preview

Abstract:

Sorbent tubes, which are packed with active carbon and Tenax-TA, used to sample the non-polar volatile organic compounds (VOCs) in ambient air, are widely accepted for its simple method, high repeatability and stability. In this paper we reported a method to prepare the standard sorbent tubes loaded with VOCs components, which is based on ISO 6145. The standard VOCs mixture gasses are generated by a dynamic volumetric diffusion or permeation method and a mass flow control pump sample system. Experiments indicate that the standard sorbent tubes have a good repeatability, linearity and stability, which are expected being used as certified reference materials in China.

You might also be interested in these eBooks

Info:

Periodical:

Advanced Materials Research (Volumes 807-809)

Pages:

118-123

Citation:

Online since:

September 2013

Authors:

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2013 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

[1] EPA, 1999. Compendium of Method for the Determination of Toxic Organic Compounds in Ambient Air, Second Edition, Compendium Method TO-17.

Google Scholar

[2] ISO 6145-8, 2005. Gas analysis ¾ Preparation of calibration gas mixtures using dynamic volumetric methods ¾ part 8: diffusion method.

DOI: 10.3403/30128055u

Google Scholar

[3] ISO 6145-10, 2002. Gas analysis ¾ Preparation of calibration gas mixtures using dynamic volumetric methods ¾ part 10: Permeation method.

DOI: 10.3403/02531857