A method was proposed for the determination of copper, iron, nickel, lead, cadmium and chromium in electroplating wastewater by ICP-AES. Working conditions and selection of spectral lines were optimized. The mixed and single standard for 6 elements were analyzed, spectrum with high sensitivity, low spectral interference, high ratio of signal with background, low detection limit were selected. The detection limit was 0.002, 0.003, 0002, 0.008, 0.002, 0.007mg/L. The effects of inhalation rate and the spectral intensity in nitric acid, hydrochloric acid, perchloric acid and sulfuric acid solution were tested. Nitrate acid with low viscosity, low surface tension and high atomization efficiency was selected. The accuracy and precision were tested, the relative error was less than 5.2%, relative standard deviation was less than 2.4%.It was accord with national standard. It proved that the determination of copper, iron, nickel, lead, cadmium and chromium in electroplating wastewater by ICP-AES was feasible.
You might also be interested in these eBooks
Info:
© 2014 Trans Tech Publications Ltd. All Rights Reserved
[1]
LU Dan, RUAN Yi. ICP-AES Determination of Lead, Cadmium, Chromium and Arsenic in Industrial Waste Water [J]. PTCA(PART B: CHEM. ANAL). Vol. 43(2007), p.1022.
Google Scholar
[2]
LI Chen, SHU Wen-bo, YANG Na-na et al. Progress in Treatment of E|ectroplating Wastewater [J]. Electroplating & Pollution Control. Vol. 30(2010), p.40.
Google Scholar
[3]
ZHOU Ju-feng, HUANG Lan-fang, PENG Li-sheng et al. Determination of Cr(VI) and Cr(III) in Electroplating Wastewater with p-amino-N, N-diethylaniline Spectrophotometric Method [J]. J. CENT. SOUTH UNIV. ( SCIENCE AND TECHNOLOGY). Vol. 36(2005).
Google Scholar
[4]
XIAO Xin-feng, LUO Ya-jan, WANG Zhao-li et a1. Reverse flow-injection spectrophotometric determination of chromium in electroplating waste water and sludge [J]. Metallurgical Analysis. Vol. 28(2008), p.47.
Google Scholar
[5]
ZHAO Xi-wen, WEN Xiang-dong. Determination of Mg, Mn, K, Na in vanadium-titanium magnetite by ICP-AES [J]. Chinese Journal of Analysis Laboratory. Vol. 24(2004), p.229.
Google Scholar
[6]
LONG Jia-hong, XU Hiong-fei, KANG Xiao-yu et a1. Determination of Cu, Ni and Sn in Wastewater of Manufactory of Circuit Board by ICP-OES [J]. Chinese Journal of Spectroscopy Laboratory. Vol. 29(2012), p.2208.
Google Scholar
[7]
YANG Xiao-yan, LI Li, WANG Bi-xuan et al. Determination of Cr in electroplating waste water by ICP-AES [J]. Materials Protection. Vol. 9(2008), p.70.
Google Scholar
[8]
GB 21900-2008 Emission Standard of pollutants for electrolating.
Google Scholar
[9]
GB/T 7475-1987Water quality; Determination of copper, zinc, lead and cadmium; Atomic absorption spectrometry.
Google Scholar
[10]
GB/T 11910-1989Water quality-Determination of nickel-Dimethylglyoxime spectrophotometric method.
Google Scholar
[11]
GB/T 11911-1989 Water quality-Determination of iron and manganese-Flame atomic absorption spectrophotometric method.
Google Scholar
[12]
GB/T 7466-1987 Water quality-Determination of total chromiun.
Google Scholar
[13]
GB/T 7470-1987 Water quality-Determination of lead-Spectrophotometric method with dithizonc.
Google Scholar