Design of the EMF-PF and Error Correction

Article Preview

Abstract:

This paper designs the electromagnetic flowmeter in partially filled pipes (EMF-PF), which sensor alternate duties, sensing both the fluid velocity and fill height signals in order to get the flow in the partially filled pipes. Because the errors caused by two parameters measured, different sensitivities and zero-shifts reduce the accuracy of the EMF-PF, therefore calibration is needed. The repeated tests show that the third-order correction of the errors achieved good results.