Numerical Investigation of the Influence of the Flow Channel Gap on Performances of Axial Flow Maglev Blood Pump

Article Preview

Abstract:

Artificial heart pump (referred to as blood pump) has become an essential means to save patients with heart failure. The pump head, the flow shear stress, blood cell flow path have a close relationship with hemodynamic characteristics. And the flow channel gap is one of the main structural parameters of the blood pump that affects the flow characteristics. For an axial flow maglev blood pump, based on CFD, using k-ε model, unstructured mesh, establishing the whole flow channel model, a three-dimensional numerical simulation of axial flow maglev blood pump was carried. The relationship between the port clearance ratio and the pump head, the flow shear stress, blood cell flow path was explored. Simulation results show that: the port clearance ratio δ=0.75, the flow channel gap h = 2.5mm, blade height b = 1.8mm, the optimal performance was attained. The pump head reached 120mmHg, the maximum shear force was less than 450Pa, both meet the functional requirements and reduce the blood damage. This results provide an important basis for structure design of axial flow maglev blood pump.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

363-368

Citation:

Online since:

August 2016

Export:

Price:

Permissions CCC:

Permissions PLS:

Сopyright:

© 2016 Trans Tech Publications Ltd. All Rights Reserved

Share:

Citation:

* - Corresponding Author

[1] Z. Qu, Modern mechanic circulation for heart failure, Scientific and Technical Documentation Press, Beijing, (2008).

Google Scholar

[2] L. H. Guo, Progress of the third generation blood pumps. Chinese J. Clinical Thoracic and Cardiovascular Surgery. 17(4) (2010) 321-325.

Google Scholar

[3] D. Xia, C. Zhao, X. Zhang, et al. Computational fluid dynamics modeling and hemolysis analysis of axial blood pumps with various impeller structures. Prog. Naturalence. 09 (2006) 993-997.

DOI: 10.1080/10020070612330100

Google Scholar

[4] Y. Peng, Y. Wu, X. Tang, et al. Numerical simulation and comparative analysis of flow field in axial blood pumps. Comput. Methods Biomech. Biomed. Eng. 17(7) (2014) 723-727.

Google Scholar

[5] Y. P. Wang, X. W. Song, C. T. Ying. Influence of diffuser for an artificial axial heart pump. Chinese J. Biomed. Eng. 26(5) (2007) 758-762.

Google Scholar

[6] G. R. Li, X. D. Zhu, Y. Y. Peng. Affection of guiding vane on axial flow pump heart assist device and its clinical relevancy. Chinese Med. Equip. J. 28(8) (2007) 5-7.

Google Scholar

[7] J. J. Yu, X. W. Zhang. Hydrodynamic and hemolysis analysis on distance and clearance between impeller and diffuser of axial blood pump. J. Mech. Med. Biol. 16(2) (2015) 1650014.

DOI: 10.1142/s0219519416500147

Google Scholar

[8] M. D. Sinnott, P. W. Cleary. Effect of rotor blade angle and clearance on blood flow through a non-pulsatile, axial, heart pump. Prog. Comput. Fluid Dyn. Int. J. 10(5/6) (2010) 300-306.

DOI: 10.1504/pcfd.2010.035363

Google Scholar